Verified computational models of total knee replacement serve as the primary design-phase tool for parametric analysis of implant geometry. Previously, dynamic finite element models of the Kansas Knee Simulator (KKS) were developed and tibiofemoral (TF) and patellofemoral (PF) kinematic predictions were verified by comparison with experimental measurements [1,2]. In this prior work, the implants were mounted in metallic fixtures to assess the ability of the model to accurately predict the TF and PF kinematics without the additional complexity of variable cadaver specimens and soft-tissue constraint. The next step in the systematic model verification procedure was to verify kinematic predictions with multiple specimen-specific models. Specifically, the objectives of the present study were: 1) to develop an explicit finite element (FE) model of the KKS capable of recreating experimental loading protocols for a deep knee bend activity and 2) to verify predicted six degree-of-freedom (DOF) TF and PF kinematics of two cruciate retaining (CR) and two posterior stabilized (PS) implanted specimen-specific models with deformable, wrapping soft tissue constraint.
Skip Nav Destination
ASME 2008 Summer Bioengineering Conference
June 25–29, 2008
Marco Island, Florida, USA
Conference Sponsors:
- Bioengineering Division
ISBN:
978-0-7918-4321-5
PROCEEDINGS PAPER
Specimen-Specific Verification of Predicted TKR Mechanics During Simulated Deep Flexion Loading Conditions
Mark A. Baldwin,
Mark A. Baldwin
University of Denver, Denver, CO
Search for other works by this author on:
Chadd W. Clary,
Chadd W. Clary
University of Kansas, Lawrence, KS
Search for other works by this author on:
Lorin P. Maletsky,
Lorin P. Maletsky
University of Kansas, Lawrence, KS
Search for other works by this author on:
Paul J. Rullkoetter
Paul J. Rullkoetter
University of Denver, Denver, CO
Search for other works by this author on:
Mark A. Baldwin
University of Denver, Denver, CO
Chadd W. Clary
University of Kansas, Lawrence, KS
Lorin P. Maletsky
University of Kansas, Lawrence, KS
Paul J. Rullkoetter
University of Denver, Denver, CO
Paper No:
SBC2008-192948, pp. 819-820; 2 pages
Published Online:
March 13, 2014
Citation
Baldwin, MA, Clary, CW, Maletsky, LP, & Rullkoetter, PJ. "Specimen-Specific Verification of Predicted TKR Mechanics During Simulated Deep Flexion Loading Conditions." Proceedings of the ASME 2008 Summer Bioengineering Conference. ASME 2008 Summer Bioengineering Conference, Parts A and B. Marco Island, Florida, USA. June 25–29, 2008. pp. 819-820. ASME. https://doi.org/10.1115/SBC2008-192948
Download citation file:
3
Views
0
Citations
Related Proceedings Papers
Related Articles
Verification of Predicted Knee Replacement Kinematics During Simulated Gait in the Kansas Knee Simulator
J Biomech Eng (August,2010)
Comparison of Deformable and Elastic Foundation Finite Element Simulations for Predicting Knee Replacement Mechanics
J Biomech Eng (October,2005)
Conceptual Design for Condylar Guiding Features of a Total Knee Replacement
J. Med. Devices (June,2011)
Related Chapters
Data Tabulations
Structural Shear Joints: Analyses, Properties and Design for Repeat Loading
List of Commercial Codes
Introduction to Finite Element, Boundary Element, and Meshless Methods: With Applications to Heat Transfer and Fluid Flow
Introduction and Scope
High Frequency Piezo-Composite Micromachined Ultrasound Transducer Array Technology for Biomedical Imaging