Stent strut configuration has been found to alter local fluid flow and wall shear stress (WSS) patterns that can promote intimal hyperplasia. To quantify these alterations, we used particle image velocimetry (PIV) to investigate local fluid dynamics distal to a deployed stent within an acrylic tube. A blood analogue mixture of glycerol and water was pumped through a flow loop in both steady and pulsatile conditions. Steady and pulsatile velocity profiles and near WSS (NWSS) predicted by PIV were in good agreement with the Poiseuille and Womersley estimations. Introduction of a Palmaz balloon expanded slotted tube stent increased centerline velocities between 6.9% and 9% and decreased NWSS distal from the stent outlet by > 40%. Peak normalized vorticity was similar between non-stented and stented flows, although spatial coverage of higher vorticity was increased upon stent introduction. Future work will incorporate Echo PIV to complement PIV findings on how strut configuration affects flow dynamics.
Skip Nav Destination
ASME 2012 Summer Bioengineering Conference
June 20–23, 2012
Fajardo, Puerto Rico, USA
Conference Sponsors:
- Bioengineering Division
ISBN:
978-0-7918-4480-9
PROCEEDINGS PAPER
The Investigation of Flow Separation Downstream of a Slotted Tube Stent in the Presence of a Pulsatile Flow Environment
Andrew M. Walker,
Andrew M. Walker
University of Calgary, Calgary, AB, Canada
Search for other works by this author on:
Clifton R. Johnston,
Clifton R. Johnston
Dalhousie University, Halifax, NS, Canada
Search for other works by this author on:
David E. Rival
David E. Rival
University of Calgary, Calgary, AB, Canada
Search for other works by this author on:
Andrew M. Walker
University of Calgary, Calgary, AB, Canada
Clifton R. Johnston
Dalhousie University, Halifax, NS, Canada
David E. Rival
University of Calgary, Calgary, AB, Canada
Paper No:
SBC2012-80290, pp. 563-564; 2 pages
Published Online:
July 19, 2013
Citation
Walker, AM, Johnston, CR, & Rival, DE. "The Investigation of Flow Separation Downstream of a Slotted Tube Stent in the Presence of a Pulsatile Flow Environment." Proceedings of the ASME 2012 Summer Bioengineering Conference. ASME 2012 Summer Bioengineering Conference, Parts A and B. Fajardo, Puerto Rico, USA. June 20–23, 2012. pp. 563-564. ASME. https://doi.org/10.1115/SBC2012-80290
Download citation file:
5
Views
0
Citations
Related Proceedings Papers
Related Articles
Two-Dimensional Optimization of a Stent for an Aneurysm
J. Med. Devices (June,2010)
Developing Pulsatile Flow in a Deployed Coronary Stent
J Biomech Eng (June,2006)
The Localized Hemodynamics of Drug-Eluting Stents Are Not Improved by the Presence of Magnetic Struts
J Biomech Eng (January,2017)
Related Chapters
Concluding remarks
Mechanical Blood Trauma in Circulatory-Assist Devices
Completing the Picture
Air Engines: The History, Science, and Reality of the Perfect Engine
Principles
Ultrasonic Methods for Measurement of Small Motion and Deformation of Biological Tissues for Assessment of Viscoelasticity