The location of High Intensity Focused Ultrasound (HIFU) beam in a tissue medium is an important parameter in assessment of the thermal field as it influences the temperature rise in the tissue. Our hypothesis is that the location of the beam can be affected by the power of the transducer. HIFU procedure with 30s of sonication time was performed at different powers of transducer (5 to 60 W) as well as different initial locations of beam in a tissue mimicking material. Eight thermocouples were embedded at 4 different layers in a phantom to measure the temperature rise during HIFU procedure. An inverse method based on experimental data and optimization algorithm was used to find the actual location of beam based on the experimental data. Our experimental data showed that for a higher power (60 W) as compared to a lower power (5 W), the focal distance that the actual position of beam moves away from its initial location increased with the raise in power. Thus, beam location can change at different powers of transducer. Using inverse method we showed that there is a direct linear correlation (R2 = 0.95) between the transducer power and the distance that beam moves away from its initial location. Therefore, it is of great clinical importance to study the effects of transducer power on the location of HIFU beam in an attempt to minimize the damage to healthy cells.
Skip Nav Destination
ASME 2012 Summer Bioengineering Conference
June 20–23, 2012
Fajardo, Puerto Rico, USA
Conference Sponsors:
- Bioengineering Division
ISBN:
978-0-7918-4480-9
PROCEEDINGS PAPER
Influence of the Transducer Acoustic Power on Focal Location of the Beam During HIFU Ablation Procedure
Seyed Ahmad Reza Dibaji,
Seyed Ahmad Reza Dibaji
University of Cincinnati, Cincinnati, OH
Search for other works by this author on:
Rupak K. Banerjee
Rupak K. Banerjee
University of Cincinnati, Cincinnati, OH
Search for other works by this author on:
Seyed Ahmad Reza Dibaji
University of Cincinnati, Cincinnati, OH
Rupak K. Banerjee
University of Cincinnati, Cincinnati, OH
Paper No:
SBC2012-80830, pp. 951-952; 2 pages
Published Online:
July 19, 2013
Citation
Dibaji, SAR, & Banerjee, RK. "Influence of the Transducer Acoustic Power on Focal Location of the Beam During HIFU Ablation Procedure." Proceedings of the ASME 2012 Summer Bioengineering Conference. ASME 2012 Summer Bioengineering Conference, Parts A and B. Fajardo, Puerto Rico, USA. June 20–23, 2012. pp. 951-952. ASME. https://doi.org/10.1115/SBC2012-80830
Download citation file:
4
Views
0
Citations
Related Proceedings Papers
Related Articles
Characterization of Focal Location During High-Intensity Focused Ultrasound Ablation in a Tissue Phantom Using Remote Thermocouple Arrays
J. Med. Devices (June,2016)
Pulsed Ultrasound Assisted Thermo-Therapy for Subsurface Tumor Ablation: A Numerical Investigation
J. Thermal Sci. Eng. Appl (August,2021)
Reduction of Noise From MR Thermometry Measurements During HIFU Characterization Procedures
J. Nanotechnol. Eng. Med (May,2011)
Related Chapters
Introduction and Scope
High Frequency Piezo-Composite Micromachined Ultrasound Transducer Array Technology for Biomedical Imaging
Tissue Harmonic Motion Estimation for Tissue Characterization Using Ultrasound and Kalman Filter
Biomedical Applications of Vibration and Acoustics in Imaging and Characterizations
Conclusions and Future Perspectives
High Frequency Piezo-Composite Micromachined Ultrasound Transducer Array Technology for Biomedical Imaging