Abstract
This paper concerns the design and rigorous in silico evaluation of a closed-loop hemorrhage resuscitation algorithm with blood pressure (BP) as controlled variable. A lumped-parameter control design model relating volume resuscitation input to blood volume (BV) and BP responses was developed and experimentally validated. Then, three alternative adaptive control algorithms were developed using the control design model: (i) model reference adaptive control (MRAC) with BP feedback, (ii) composite adaptive control (CAC) with BP feedback, and (iii) CAC with BV and BP feedback. To the best of our knowledge, this is the first work to demonstrate model-based control design for hemorrhage resuscitation with readily available BP as feedback. The efficacy of these closed-loop control algorithms was comparatively evaluated as well as compared with an empiric expert knowledge-based algorithm based on 100 realistic virtual patients created using a well-established physiological model of cardiovascular (CV) hemodynamics. The in silico evaluation results suggested that the adaptive control algorithms outperformed the knowledge-based algorithm in terms of both accuracy and robustness in BP set point tracking: the average median performance error (MDPE) and median absolute performance error (MDAPE) were significantly smaller by >99% and >91%, and as well, their interindividual variability was significantly smaller by >88% and >94%. Pending in vivo evaluation, model-based control design may advance the medical autonomy in closed-loop hemorrhage resuscitation.
Introduction
Hemorrhage is accountable for approximately 40% of trauma mortality worldwide [1]. Hemorrhage is known to be responsible for the majority of preventable deaths due to traumatic injuries that occur before arrival to receive definitive care at hospitals. It is predicted that approximately 25% of mortality due to hemorrhage may be preventable with early resuscitation and intervention [2]. Hence, early life-saving interventions to treat hemorrhage are a cornerstone to reduce trauma-induced mortality. In fact, shorter prehospital time has been shown to improve survival after traumatic injuries with hemorrhage [3]. However, rapid transport to a hospital equipped with definitive care capability is not always possible. From this standpoint, advanced hemorrhage control capability compatible with prehospital and austere environments may have the potential to improve the mortality rate of hemorrhaging patients.
Hemorrhage is resuscitated in the field by stopping bleeding if possible and then administering volume (i.e., fluids) to the patient to compensate for the volume deficit due to blood loss. In contemporary clinical practice, hemorrhage resuscitation is predominantly manual volume boluses administration by clinicians to a physiological endpoint, e.g., blood pressure (BP) [4]. Such a manual administration of volume can be ad hoc and subjective due to its dependence on individual clinician's experience and knowledge [5–10] as well as their lapses of vigilance and suboptimal decision-making [11–13]. Noting that the therapeutic range of hemorrhage resuscitation is narrow and the response to hemorrhage resuscitation exhibits a large interindividual variability [14,15], novel capabilities to precisely administer volume (both in terms of timing and dose) are essential in order to drastically advance hemorrhage resuscitation by addressing the abovementioned shortcomings.
A potential solution is to computerize and automate the hemorrhage resuscitation task. Indeed, computerized autonomy can enable dedicated monitoring and treatment of patients with no lapses of vigilance, especially if it is equipped with trustworthy decision-making algorithms. Pioneering work to leverage computational autonomy has been reported in the context of closed-loop control of hemorrhage resuscitation [16–22]. These reports strongly suggest the promise of computerized closed-loop administration of hemorrhage resuscitation therapy. Yet, most, if not all, of the existing closed-loop control algorithms have been developed by translating qualitative expert knowledge into empiric rules, including (i) population statistics combined with adaptive error correction and an empiric rule base [17,18,20]; (ii) decision table [16,19]; (iii) proportional–integral–derivative [23,24]; and (iv) fuzzy logic [21,23]. Such rule-based algorithms were shown to be efficacious in pilot in vivo animal and human trials. However, regulatory certification and clinical use of these algorithms require strict understanding of their safety and robustness characteristics as well as limitations, which is usually derived from mathematical analysis in the context of control theory. From this standpoint, qualitative, complex, and nonlinear nature of these rule-based control algorithms, in conjunction with the lack of physiological models for stability and robustness analysis of closed-loop control algorithms, may present challenges in conducting such an analysis. In fact, state-of-the-art physiological models involve excessive details, such as ion and protein kinetics, which introduce formidable complexities, making the physiological models less ideal for control design and analysis [25–34]. An alternative approach to overcome this obstacle may be to develop a physiological model amenable to control design (called control design model) and to design a closed-loop control algorithm systematically by exploiting the control design model. This is in fact a well-established procedure in the domain of control systems engineering. Yet, it has not been adopted for the development of closed-loop hemorrhage resuscitation algorithms, because no physiological model appropriate for use in the design and analysis of closed-loop fluid resuscitation controllers exists to the best of our knowledge.
This paper concerns the design of a closed-loop hemorrhage resuscitation algorithm with BP as controlled variable. To the best of our knowledge, this is the first work to demonstrate model-based control design for hemorrhage resuscitation with readily available BP as feedback. A lumped-parameter control design model relating volume resuscitation input to blood volume (BV) and BP responses was developed and experimentally validated. Then, three alternative adaptive control algorithms were developed using the control design model for hemorrhage resuscitation: (i) model reference adaptive control (MRAC) with BP feedback, (ii) composite adaptive control (CAC) with BP feedback, and (iii) composite adaptive control with BV and BP feedback. The efficacy of these closed-loop hemorrhage resuscitation control algorithms was evaluated as well as compared with an empiric expert knowledge-based algorithm based on 100 realistic virtual patients created using a well-established physiological model of cardiovascular (CV) hemodynamics.
Control Design Model
Model Description.
We developed a lumped-parameter control design model representing the dynamical relationship between resuscitation volume versus BV and BP responses (Fig. 1). The control design model is built upon a lumped-parameter model of BV kinetics [35] and extends it to also include the dynamical relationship between BV and BP with a time-varying gain representing the resultant CV actions on BV to elicit BP.
In this way, the model structure in Eq. (8) may be well supported by the physical principles underlying volume resuscitation, including (i) the conservation of volume (Eq. (1)), (ii) intercompartmental fluid exchange (Eq. (3)), and (iii) time-varying BV–BP relationship (Eq. (7)), under modest and perhaps justifiable assumptions.
In this work, we will consider BV and BP as measured variables. The rationale underlying BP is its ease of continuous clinical measurement (e.g., with an arterial catheter) despite its complex relationship with BV, while the rationale underlying BV is its potential as an emerging modality for critical care. Indeed, commercialized medical devices have recently appeared in the market to continuously measure blood hematocrit (HCT) [41], based on which BV can be estimated [42].
Model Validation: Methods.
We examined the validity of the control design model using experimental data collected from 23 sheep undergoing BV perturbations in a prior work [16]. Each animal was subject to (i) controlled hemorrhage of 25 ml/kg over 15 min, (ii) 150-min crystalloid resuscitation starting 30 min after the hemorrhage (the rate of which was determined by a previously investigated closed-loop control algorithm aimed at regulating BP in the sheep [16]), and (iii) controlled hemorrhage of 5 ml/kg over 5 min at 35 min and 55 min after the initial hemorrhage. Measurements made include rates of hemorrhage and urinary output () and crystalloid infusion (), as well as BV (using HCT measurements in conjunction with the indocyanine green dye), BP, and cardiac output.
We validated the control design model on an individual basis. Given that Eq. (8) is a time-varying equation due to , conventional batch system identification cannot be applied in fitting Eq. (8) to the experimental data. Existing remedies to cope with time-varying parameters (e.g., repetitive generalized least squares across multiple small time intervals followed by averaging [43]) are not suited to our purpose, because they cannot strictly capture the instantaneous value of at all time instants. Hence, we formulated a novel two-layer optimization problem in Fig. 3. In the upper layer, the parameters , , and in Eq. (8) are optimized, so that Eq. (8) can best fit the experimental data (BV and BP in particular) in the root-mean-squared sense given
where and are the changes in BV and BP predicted by the control design model, is the actual change in BP, is the adaptation gain, and . We used the experimental data as the initial conditions for BV () and BP () in solving Eqs. (4), (7), and (10). Hence, and . These two layers are iteratively executed until the parameter estimates converge. We implemented and solved the two-layer optimization problem using matlab and its optimizationtoolbox (MathWorks, Natick, MA).
Model Validation: Results and Discussion.
Table 1 summarizes the root-mean-squared fitting errors associated with BV, BP, and (where ground truth was computed as from direct measurements of BV () and BP () at each time ), both in the absolute and percentage sense, as well as the values of the parameter estimates . Figure 4 shows a representative example of the results obtained from the two-layer optimization in fitting the control design model to experimental data in a sheep. Once fitted to each sheep, the control design model could accurately replicate the experimental data on BV and BP responses to both hemorrhage and crystalloid resuscitation. In addition, the control design model could track the time-varying parameter . It is noted that was estimated using (i.e., predicted by Eq. (4) but not its measurement) and . Hence, accurate tracking of experimental shown in Fig. 4 is not trivial. The ability of the control design model to closely reproduce experimental data suggests that it is appropriate to the design of closed-loop control algorithms for hemorrhage resuscitation. In addition, the validation results also suggest that adaptive control may be a reasonable approach to closed-loop control of hemorrhage resuscitation: (i) the intraindividual variability in can be adequately tracked with a recursive estimation compatible with adaptive control algorithms, and (ii) the interindividual variability in the parameter estimates is substantial (45% for and 67% for , in terms of the coefficient of variation).
(a) Root-mean-squared fitting errors associated with BV, BP, and in the absolute and percentage sense | ||
---|---|---|
BV | BP | |
92 ± 39 (ml) | 4.7 ± 2.8 (mm Hg) | 9.6 × 10−3 ± 4.7 × 10−3 (mm Hg/ml) |
(12.8 ± 4.9%) | (8.6 ± 4.7%) | (7.7 ± 3.2%) |
(a) Root-mean-squared fitting errors associated with BV, BP, and in the absolute and percentage sense | ||
---|---|---|
BV | BP | |
92 ± 39 (ml) | 4.7 ± 2.8 (mm Hg) | 9.6 × 10−3 ± 4.7 × 10−3 (mm Hg/ml) |
(12.8 ± 4.9%) | (8.6 ± 4.7%) | (7.7 ± 3.2%) |
(b) The values of the parameter estimates | ||
---|---|---|
(min−1) | ||
3.3 ± 1.3 | 1.3 ± 0.8 | 0.4 ± 0.2 |
(b) The values of the parameter estimates | ||
---|---|---|
(min−1) | ||
3.3 ± 1.3 | 1.3 ± 0.8 | 0.4 ± 0.2 |
Control Design
The large interindividual parametric variability in the control design model (shown in Control Design Model section) naturally motivates us to explore adaptive control in the design of closed-loop control algorithms for hemorrhage resuscitation. To enable the application of adaptive control theory to hemorrhage resuscitation problem with the plant dynamics given by Eq. (8), we make the following assumptions:
The parameter is slowly varying and may be regarded as constant for the sake of control design: . That is not strictly true due to the autonomic-cardiac regulation actions in response to volume resuscitation. However, recursive parameter estimation may still be able to track the time-varying changes in as shown in Fig. 4.
The sign of is known. is the high-frequency gain of the control design model in Eq. (8). In the physiological context, the sign of depends on and (see Eq. (7)). Experimental data suggest that while on the average. Hence, it is not straightforward to define the sign of . However, it is well known that the relationship between BV and BP during volume resuscitation is passive (i.e., proportional with positive sign) in the absence of hemorrhage [39]: resuscitation volume increases BV, and BP also increases as BV increases. Hence, may be safely assumed. A potential drawback is that it restricts the context of use of the closed-loop control algorithm to hemorrhage resuscitation scenarios where blood loss is stopped before resuscitation starts.
In the adaptive control paradigm, the unknown parameters in the plant dynamics in Eq. (8) can be recursively estimated using the input–output measurements. However, accurate estimation of all these unknown parameters requires the persistence of excitation condition [44,45], which is not easy to anticipate in real-world hemorrhage resuscitation scenarios (simply because the goal of the resuscitation is to save the patient rather than to meet the persistence of exictation condition). In the lack of persistence of excitation in the resuscitation input , the parameter estimates obtained from recursive estimation may suffer from inaccuracy. To overcome this challenge, we treat as the only unknown parameter in the control design model, while is known a priori. This assumption is justifiable in that (i) current practice of hemorrhage resuscitation consists of iterative administration of multiple boluses, and (ii) may be readily estimated even using the measurements of the bolus input and BP output associated with a single bolus administration by performing batch system identification of the control design model in Eq. (8) (see In Silico Evaluation: Results and Discussion section for details) [35].
To achieve the control objective of regulating BP at a desired level, we comparatively investigate three alternatives: (i) MRAC with BP feedback alone, (ii) CAC with BP feedback alone, and (iii) CAC with BP and BV feedback.
Model Reference Adaptive Control With Blood Pressure Feedback.
where is the differentiation operator.
yields negative semidefiniteness of : . Consequently, is bounded. Hence, the closed-loop dynamics consisting of Eqs. (11), (13), and (17) is globally stable. In turn, the signals , , and are bounded, and accordingly in Eq. (14) is also bounded. As a result, is bounded, rendering uniformly continuous. By invoking Barbalat's lemma, it is concluded that . Hence, we can conclude that the tracking error converges to zero.
Composite Adaptive Control With Blood Pressure Feedback.
Hence, the argument made on the convergence of the tracking error remains valid for CAC with BP feedback.
Composite Adaptive Control With Blood Pressure and Hematocrit Feedback.
To examine the potential value of direct real-time BV measurement (which is not widely used now but may be readily available in the near future with real-time HCT monitoring capability [41]) in addition to BP, we explored the use of CAC based on BV feedback, in which the adaptation law in Eq. (17) (which is based on the tracking error ) is augmented by an extra term governed by a prediction error associated with both BV and BP.
Hence, the argument made on the convergence of the tracking error remains valid for CAC with BV and BP feedback.
Control Algorithm Evaluation
In Silico Patient Model and Virtual Patient Generation.
To evaluate the alternative closed-loop hemorrhage resuscitation control algorithms developed in Control Design section, we used an established and comprehensive physiological model of CV hemodynamics in humans [46]. The physiological model consists of essential components to faithfully describe CV hemodynamics, including: (i) circulatory dynamics in the arteries and the veins, (ii) interaction between intravascular and extravascular compartments, (iii) renal function, (iv) sympathetic and parasympathetic feedback, (v) peripheral autoregulation, and (vi) renin–angiotensin control mechanism (Fig. 5). Hence, the physiological model is a powerful tool for simulating short-term and long-term regulation of BV and BP. More specifically, the circulatory dynamics represents the BV dynamics in the intravascular compartment that includes arteries, veins, and the heart as compliant chambers holding the blood. BV and BP are regulated by a suite of autonomic regulation mechanisms including the intercompartmental fluid exchange between intravascular (i.e., blood) and extravascular compartments, the renal function in the kidneys (i.e., urine output), and many BP control functions (such as the autonomic control of cardiac contractility, vascular resistance, and unstressed venous BV, autoregulation as a function of capillary blood flow, and angiotensin control).
where is RBCV, and . Then, plasma volume () can be computed by .
The physiological model is equipped with >40 parameters characterizing the interactions therein to control BP. We selected nine parameters whose variability largely alters BP, including: (i) nominal arterial resistance, (ii) nominal cardiac contractility, (iii) nominal capillary filtration, (iv) nominal venous capacitance, (v) the effect of sympathetic tone on kidney function, (vi) nominal unstressed and total venous BV, and (vii) parameters to characterize angiotensin function. Then, we created a total of 100 virtual patients by randomly sampling these parameters from physiologically reasonable ranges in the vicinity of their respective nominal values, so that BV and BP associated with the virtual patients in the prehemorrhagic state exhibit physiologically adequate values. The resulting virtual patients were associated with prehemorrhage BV range of 4.5–5.5 l and BP range of 70–100 mm Hg.
In Silico Evaluation: Methods and Analysis.
To evaluate the closed-loop control algorithms developed in Control Design section, we devised a hemorrhage and resuscitation scenario shown in Fig. 6. In this scenario, the virtual patient is subject to major hemorrhage (2.5 l). The hemorrhage is treated during the next 120 min. Then, a 0.5 l bolus is administered during 30 min. The closed-loop control algorithm is activated 10 min after the bolus is administrated. We executed this scenario in all the 100 virtual patients. Then, we collectively analyzed the efficacy of the closed-loop control algorithms by aggregating the in silico evaluation results obtained from all the virtual patients.
where is HCT. Then, each closed-loop control algorithm was implemented using the derived parameter estimates and then executed to achieve the objective of regulating BP at a target set point (which in this work was set at 100 mm Hg in all virtual patients). For all the closed-loop control algorithms, we set the time constant associated with the reference model ( in Eq. (12)) at 15 min.
To compare the efficacy of our closed-loop control algorithms with a benchmark, we evaluated an expert knowledge-based closed-loop hemorrhage resuscitation algorithm reported in a prior work [16,21,47], slightly revised to achieve our 100 mm Hg set point. In brief, the expert knowledge-based algorithm delivers volume at prespecified BP-dependent rates: 100 ml/min per 70 kg body weight if BP < 40 mm Hg, its 80% level if BP is 41–44 mm Hg, 60% level if BP is 45–49 mm Hg, 30% level if BP is 50–69 mm Hg, and 10% level if BP is 70–99 mm Hg. If BP is 100 mm Hg or above, no volume was delivered.
To make the in silico scenario close to the reality, we added measurement noises to BP and HCT signals: BP signal was contaminated with a uniform random noise of 2 mm Hg in magnitude, while HCT signal was contaminated with a uniform random noise of 0.01 in magnitude. The closed-loop control algorithms were equipped with a six-point moving average digital filter to reject the adverse impact of noises. In reality, a BP transducer and a volumetric infusion pump are needed to implement the control loop. Yet, considering that the time constants associated with the BP transducer and the infusion pump are much faster than that of the BV kinetics, dynamic effect of these physical elements was not considered.
Using these PEs, we evaluated the efficacy of the closed-loop control algorithms by quantifying, assessing, and comparing the median PE (MDPE), median absolute PE (MDAPE), divergence (DIV, which is a measure of drift in PE), and wobble (which is a measure of intraindividual variability in PE). We used the one-way analysis of variance (ANOVA) with post hoc paired t-test to determine the statistical significance in the difference in these measures between the closed-loop control algorithms developed in Control Design section. We used the paired t-test to determine the statistical significance in the difference in these measures between CAC with BV and BP feedback (Eqs. (13) and (27)) and the benchmark expert knowledge-based hemorrhage resuscitation algorithm.
In Silico Evaluation: Results and Discussion.
Figure 7 presents the resuscitation input (, first row), BV (second row), BP (third row), and finally and (fourth row) trajectories when (a) MRAC with BP feedback, (b) CAC with BP feedback, and (c) CAC with BV and BP feedback were applied to the nominal virtual patient (BV: 5.0 l and BP: 100 mm Hg) subject to the scenario shown in Fig. 6. Table 2 summarizes the PE-based metrics associated with set point tracking and parameter estimation pertaining to all the closed-loop control algorithms (it is noted that the PE-based metrics were computed from the onset of closed-loop control (t = 280 min in Fig. 6) to t = 360 min when BP reached the set point). Figure 8 presents the resuscitation input (, first column), BV (second column), and BP (third column) when the expert knowledge-based hemorrhage resuscitation algorithm was applied to the nominal virtual patient (BV: 5.0 l and BP: 100 mm Hg) subject to the scenario shown in Fig. 6, while Table 3 summarizes the PE-based metrics associated with set point tracking pertaining to the expert knowledge-based hemorrhage resuscitation algorithm.
(a) Set point tracking | ||||
---|---|---|---|---|
MDPE (%) | MDAPE (%) | DIV (%/min) | Wobble (%) | |
MRAC with BP | −0.29 ± 2.51 | 3.32 ± 0.94 | −0.15 ± 0.10 | 3.21 ± 1.06 |
CAC with BP | −0.44 ± 2.18 | 3.35 ± 0.88 | −0.19 ± 0.15a | 3.45 ± 1.23 |
CAC with BV and BP | −0.45 ± 1.14 | 2.08 ± 0.58a,b | −0.14 ± 0.07b | 2.02 ± 0.57a,b |
(a) Set point tracking | ||||
---|---|---|---|---|
MDPE (%) | MDAPE (%) | DIV (%/min) | Wobble (%) | |
MRAC with BP | −0.29 ± 2.51 | 3.32 ± 0.94 | −0.15 ± 0.10 | 3.21 ± 1.06 |
CAC with BP | −0.44 ± 2.18 | 3.35 ± 0.88 | −0.19 ± 0.15a | 3.45 ± 1.23 |
CAC with BV and BP | −0.45 ± 1.14 | 2.08 ± 0.58a,b | −0.14 ± 0.07b | 2.02 ± 0.57a,b |
(b) Parameter estimation | ||||
---|---|---|---|---|
MDPE (%) | MDAPE (%) | DIV (%/min) | Wobble (%) | |
MRAC with BP | −25.50 ± 19.35 | 27.70 ± 16.01 | −0.05 ± 0.13 | 1.42 ± 0.61 |
CAC with BP | −35.92 ± 11.13a | 35.93 ± 11.13a | 0.11 ± 0.14a | 1.58 ± 0.49a |
CAC with BV and BP | −1.44 ± 6.38a,b | 5.54 ± 3.44a,b | −0.03 ± 0.02b | 0.00 ± 0.01a,b |
(b) Parameter estimation | ||||
---|---|---|---|---|
MDPE (%) | MDAPE (%) | DIV (%/min) | Wobble (%) | |
MRAC with BP | −25.50 ± 19.35 | 27.70 ± 16.01 | −0.05 ± 0.13 | 1.42 ± 0.61 |
CAC with BP | −35.92 ± 11.13a | 35.93 ± 11.13a | 0.11 ± 0.14a | 1.58 ± 0.49a |
CAC with BV and BP | −1.44 ± 6.38a,b | 5.54 ± 3.44a,b | −0.03 ± 0.02b | 0.00 ± 0.01a,b |
Significantly different from MRAC with BP (one-way ANOVA).
Significantly different from CAC with BP (one-way ANOVA).
All the closed-loop hemorrhage resuscitation algorithms developed in this work performed well in tracking the BP set point (Fig. 7 and Table 2). Regarding the BP set point tracking error, all three closed-loop control algorithms consistently exhibited small bias, inaccuracy, drift, and fluctuations as indicated by small MDPE, MDAPE, divergence, and wobble, respectively, across all the virtual patients. Comparing the three closed-loop control algorithms, CAC with BV and BP feedback was generally superior to MRAC and CAC with BP feedback in all the PE-based metrics. Specifically, the former was superior in terms of MDAPE, divergence, and wobble. In addition, all the metrics associated with the former were more robust in that the spread of the metrics was relatively narrow. However, the difference in these metrics (and accordingly, the overall set point tracking performance) was deemed small from practical perspective. In contrast, regarding the parameter estimation error, CAC with BV and BP feedback exhibited significantly superior performance to MRAC and CAC with BP feedback in all the PE-based metrics (Fig. 7 and Table 2). Indeed, all the PE-based metrics associated with the former were close to zero, whereas the same metrics associated with the latter clearly indicated large bias (negative MDPE, meaning overestimation), poor accuracy (MDAPE being the same as MDPE except sign, meaning again consistently biased overestimation), lack of error convergence (negligible drift with close-to-zero divergence, meaning persistent parameter estimation error), and fluctuation (wobble larger than the former, although small in absolute sense). The notable superiority in parameter estimation performance associated with CAC with BV and BP feedback may be largely attributed to its exploitation of BV measurement. Indeed, the results of bolus-based system identification (Eqs. (30) and (31)) showed that more accurate estimates of the control design model parameters resulted when BV was used in conjunction with BP than when only BP was used: on the average, the errors associated with , , and were 56% and 31%, 176% and 9%, and 12% and 12%, respectively, for Eqs. (30) and (31). In addition, the error associated with for Eq. (31) was generally small (2% on the average). During closed-loop control, the integrity of is largely affected by the accuracy of and (in all three closed-loop control algorithms) as well as (in CAC with BV and BP feedback), since these parameters are not updated but fixed in implementing the control law in Eq. (13). Considering that the errors associated with these parameters are relatively large in MRAC and CAC with BP feedback than in CAC with BV and BP feedback, derived from the adaptation laws in Eqs. (17) and (21) was forced to drift away from its actual value in order to compensate for the inaccuracy in and . On the contrary, derived from the adaptation law in Eq. (27) was much more accurate by virtue of the superior accuracy in , , and . All in all, these results are consistent with the intuitive expectation that supplying more information to adaptive control can enable enhanced set point tracking (despite modest extent in case of this work) and parameter estimation.
In case was not adapted during closed-loop control (i.e., after “batch system identification” in Fig. 6), set point tracking performance was degraded, especially in terms of robustness, relative to CAC with BV and BP feedback: (i) all the metrics exhibited much larger spread, and (ii) MDAPE and wobble were significantly larger (results not shown). The results demonstrate the advantage of accurate tracking of in achieving robust set point tracking performance. It is noted that the PE metrics obtained without adapting were not significantly different from those obtained for MRAC and CAC with BP feedback, which may be attributed to, among other reasons, (i) relatively small changes in in the in silico evaluation conducted in this work (see Fig. 6) and (ii) inaccurate estimation of when only BP was used as feedback. However, we anticipate that MRAC and CAC only with BP feedback may still be advantageous in patients undergoing large time-varying changes in (which needs to be collaborated in a follow-up work).
Compared with the closed-loop control algorithms developed in this work, the expert knowledge-based hemorrhage resuscitation algorithm exhibited very poor set point tracking performance (Fig. 8). In particular, BP response was very slow: it required additional >80 min to reach the set point on the average even with an initial bolus (VF in Fig. 6). In the absence of the initial bolus (note that the expert knowledge-based hemorrhage resuscitation algorithm as reported in the prior work is not supposed to require initial bolus), it required additional >140 min to reach the set point on the average (not shown). Accordingly, all the PE-based metrics except wobble were substantially deteriorated compared with the closed-loop control algorithms developed in this work (Table 3): (i) MDPE and MDAPE were both very large, indicating slow BP response; and (ii) divergence was persistently positive, indicating gradual convergence of BP to its set point. Wobble on the average was smaller in the expert knowledge-based algorithm than in all the closed-loop control algorithms developed in this work, because the volume delivery rate was always constant (at 10 ml/min per 70 kg body weight, since BP was always between 70 and 99 mm Hg in all the virtual patients up to t = 360 min in Fig. 8). However, the expert knowledge-based hemorrhage resuscitation algorithm behaved like an on–off control once BP reached its set point, switching between 0 ml/min and 10 ml/min (both per 70 kg body weight) susceptibly to the measurement noise associated with BP (not shown). All in all, the slow response associated with the expert knowledge-based algorithm may be enhanced by adjusting its volume delivery rates. However, such enhancement may still necessitate tedious trial-and-error and ad hoc tuning. Furthermore, even the enhanced expert knowledge-based algorithm may not perform as efficaciously as any of the closed-loop control algorithms developed in this work on an individual basis, due to its population-average nature. In this regard, the control design model presented in this work (see Control Design Model section) provides a solid foundation for systematically designing closed-loop control algorithms capable of personalized hemorrhage resuscitation by tailoring its actions to individual patients via online adaptation.
Conclusion
In this paper, we developed a control-oriented model applicable to systematic design of closed-loop control algorithms for hemorrhage resuscitation and demonstrated its use in the design of three alternative adaptive control algorithms. We experimentally showed that the control design model can replicate BV and BP responses to hemorrhage and hemorrhage resuscitation, thus making it suited to the design of closed-loop control algorithms exploiting BV and BP as controlled variables. Based on an extensive in silico evaluation, we showed that our adaptive control algorithms were significantly superior to an existing expert knowledge-based hemorrhage resuscitation algorithm. Given that the context of use of our closed-loop control algorithms is limited to the scenarios in which hemorrhage has been treated, future work must investigate the development of closed-loop control algorithms equipped with extended context of use.
Acknowledgment
Any opinions, findings, and conclusions or recommendations expressed in this material are those of the authors and do not necessarily reflect the views of the NSF and ONR.