Muscles actuating the shoulder girdle are important for stabilizing the scapula and coordinating phased kinematics of the shoulder complex. If these muscles become weak or imbalanced, joint instability and injury may result. Reliable measurement of shoulder strength is thus important for prevention, diagnosis, and rehabilitation of shoulder problems. To date, studies quantifying the strength of the shoulder girdle are limited. The purpose of this work was to design and evaluate a custom apparatus and corresponding protocol for measuring maximal, voluntary, isometric strength of the shoulder girdle during various forms of shrugging exercise. A custom apparatus was constructed as a rigid frame with a vertical post supporting a seat, seat back, and horizontal beam. The beam extends laterally on either side beyond and around the shoulders of a seated subject. A pair of arm extension members pivots on the beam about an axis aligned with the shoulder flexion-extension axis. These members can be locked in place at any angle. Between them is mounted a force-sensing grip assembly, which can be adjusted proximally or distally to accommodate varying shoulder girdle positions. Subjects grasp the grip assembly handles with extended elbows and push or pull as forcefully as possible. Nine female and ten male subjects participated in a protocol using the apparatus to measure maximum isometric force generated at three positions each for elevation, depression, protraction, and retraction of the shoulder girdle . A video motion capture system was used to measure shoulder girdle angles. The reliability of the approach was evaluated based on the repeatability of measured shoulder elevation angle, protraction angle, and total force over three days of testing. The apparatus performed well during the tests, providing a stable, rigid, yet adjustable platform for measuring shoulder girdle strength. Repeatability of force measurements was interpreted as very good to excellent, with intraclass correlation coefficient (ICC) (2,1) values ranging from 0.83 to 0.95 for all tests except one . Repeatability of angle measurements was interpreted as good to excellent. For tests measuring elevation and depression strength, the ICC of elevation angle ranged from 0.85 to 0.89. For tests measuring protraction and retraction strength, the ICC of protraction angle ranged from 0.68 to 0.88. This type of apparatus could be an effective clinical tool for measuring strength in the shoulder girdle muscles. Use of the video motion capture system is optional.
Skip Nav Destination
briaṉgarner@baylor.edu
Article navigation
December 2007
Research Papers
An Apparatus and Protocol to Measure Shoulder Girdle Strength
Brian A. Garner, Ph.D.,
Brian A. Garner, Ph.D.
Department of Mechanical Engineering,
briaṉgarner@baylor.edu
Baylor University
, One Bear Place 97356, Waco, Texas 76798
Search for other works by this author on:
Jaeho Shim, Ph.D.,
Jaeho Shim, Ph.D.
Department of Health, Human Performance, and Recreation,
Baylor University
, One Bear Place 97313, Waco, Texas 76798
Search for other works by this author on:
Scott Wilson, BS
Scott Wilson, BS
Department of Mechanical Engineering,
Baylor University
, One Bear Place 97356, Waco, Texas 76798
Search for other works by this author on:
Brian A. Garner, Ph.D.
Department of Mechanical Engineering,
Baylor University
, One Bear Place 97356, Waco, Texas 76798briaṉgarner@baylor.edu
Jaeho Shim, Ph.D.
Department of Health, Human Performance, and Recreation,
Baylor University
, One Bear Place 97313, Waco, Texas 76798
Scott Wilson, BS
Department of Mechanical Engineering,
Baylor University
, One Bear Place 97356, Waco, Texas 76798J. Med. Devices. Dec 2007, 1(4): 246-253 (8 pages)
Published Online: August 10, 2007
Article history
Received:
August 25, 2006
Revised:
August 10, 2007
Citation
Garner, B. A., Shim, J., and Wilson, S. (August 10, 2007). "An Apparatus and Protocol to Measure Shoulder Girdle Strength." ASME. J. Med. Devices. December 2007; 1(4): 246–253. https://doi.org/10.1115/1.2796190
Download citation file:
25
Views
0
Citations
Get Email Alerts
Cited By
Comanipulation Robotic Platform for Spine Surgery With Exteroceptive Visual Coupling: Development and Experimentation
J. Med. Devices (December 2022)
Preclinical Evaluation of a Task-Shifting Contraceptive Implant Insertion Device for Use in Low- and Middle-Income Countries
J. Med. Devices (December 2022)
Related Articles
A Measurement Device for Rotational Stiffness of the Eye
J. Med. Devices (June,2008)
Computational Modeling to Predict Mechanical Function of Joints: Application to the Lower Leg With Simulation of Two Cadaver Studies
J Biomech Eng (December,2007)
Designing an Optical Bendloss Sensor for Clinical ForcMeasurement
J. Med. Devices (June,2009)
Measurements of the Static Friction Coefficient Between Bone and Muscle Tissues
J Biomech Eng (August,2010)
Related Proceedings Papers
Related Chapters
Intuitive Optimization
Engineering Optimization: Applications, Methods, and Analysis
Rationale for Human-Powered Vehicle Design and Use
Design of Human Powered Vehicles
Health and Safety and Emergency Response
Pipeline Transportation of Carbon Dioxide Containing Impurities