The application of stem surface treatments and finishes are common methods for improving stem-cement interface stability in joint replacement systems; however, success of these surfaces has been variable. As opposed to applying a treatment or finish, altering stem design through changing the surface topography of the base stem material may offer some advantages. This study compared the effect of stem circumferential grooving on the torsional and axial stability of cemented stems. Fifteen metal stems were machined from cobalt chrome to have smooth (n = 5) or circumferential-grooved surfaces, where groove depth and spacing was either 0.6 mm (n = 5) or 1.1 mm (n = 5). Stems were potted in aluminum tubes using bone cement, left 24 h to cure, and placed in a materials testing machine for testing using a cyclic staircase loading protocol at 1.5 Hz. All stems were tested independently in compression and torsion on separate testing days, using the same stems repotted with new cement. Motion of the stem was tracked, and failure was defined either as rapid increase in stem motion, or completion of the loading protocol. Statistical analysis was used to compare interface strength and stem motion prior to failure. Grooved stems demonstrated increased interface strength (p < 0.001) and reduced motion (p < 0.01) compared to smooth stems under compression. In torsion, no significant difference was found in strength among the grooved and smooth stems (p = 0.10); however, grooved 1.1 mm demonstrated greatest interface motion prior to catastrophic failure (p < 0.01). Overall, circumferential-grooved stems offered improved stability under compression, and comparable stability in torsion, relative to the smooth stems.
Skip Nav Destination
Article navigation
March 2014
Technical Briefs
The Effect of Stem Circumferential Grooves on the Stability at the Implant-Cement Interface
Yara K. Hosein,
Yara K. Hosein
Biomedical Engineering Graduate Program,
Western University
,London, ON N6A 5B9
, Canada
Search for other works by this author on:
Graham J. W. King,
Graham J. W. King
Department of Surgery,
Western University
,London, ON N6A 5B9
, Canada
Search for other works by this author on:
Cynthia E. Dunning
Cynthia E. Dunning
1
Biomedical Engineering Graduate Program,
Department of Surgery, and
Department of Materials & Mechanical Engineering,
e-mail: cdunning@uwo.ca
Department of Surgery, and
Department of Materials & Mechanical Engineering,
Western University
,London, ON N6A 5B9
, Canada
e-mail: cdunning@uwo.ca
1Corresponding author.
Search for other works by this author on:
Yara K. Hosein
Biomedical Engineering Graduate Program,
Western University
,London, ON N6A 5B9
, Canada
Graham J. W. King
Department of Surgery,
Western University
,London, ON N6A 5B9
, Canada
Cynthia E. Dunning
Biomedical Engineering Graduate Program,
Department of Surgery, and
Department of Materials & Mechanical Engineering,
e-mail: cdunning@uwo.ca
Department of Surgery, and
Department of Materials & Mechanical Engineering,
Western University
,London, ON N6A 5B9
, Canada
e-mail: cdunning@uwo.ca
1Corresponding author.
Manuscript received March 5, 2013; final manuscript received September 16, 2013; published online December 6, 2013. Assoc. Editor: Venketesh N. Dubey.
J. Med. Devices. Mar 2014, 8(1): 014504 (5 pages)
Published Online: December 6, 2013
Article history
Received:
March 5, 2013
Revision Received:
September 16, 2013
Citation
Hosein, Y. K., King, G. J. W., and Dunning, C. E. (December 6, 2013). "The Effect of Stem Circumferential Grooves on the Stability at the Implant-Cement Interface." ASME. J. Med. Devices. March 2014; 8(1): 014504. https://doi.org/10.1115/1.4025468
Download citation file:
Get Email Alerts
Cited By
Comanipulation Robotic Platform for Spine Surgery With Exteroceptive Visual Coupling: Development and Experimentation
J. Med. Devices (December 2022)
Preclinical Evaluation of a Task-Shifting Contraceptive Implant Insertion Device for Use in Low- and Middle-Income Countries
J. Med. Devices (December 2022)
Related Articles
The Effect of Cross-Sectional Stem Shape on the Torsional Stability of Cemented Implant Components
J Biomech Eng (June,2007)
Tensile Fatigue Failure of Acrylic Bone Cement
J Biomech Eng (November,1983)
Design of an Endoreactor for the Cultivation of a Joint-Like-Structure
J. Med. Devices (June,2009)
Design Optimization of a Prosthesis Stem Reinforcing Shell in Total Hip Arthroplasty
J Biomech Eng (August,1990)
Related Proceedings Papers
Related Chapters
Some Recent Advances and Current Challenges in Joint-Replacement Tribology and UHMWPE Bearing Surfaces
Tribology of Mechanical Systems: A Guide to Present and Future Technologies
Basic Concepts
Design & Analysis of ASME Boiler and Pressure Vessel Components in the Creep Range
Surface Topography Simulation Technology for Vibration-Assisted Machining
Vibration Assisted Machining: Theory, Modelling and Applications